PROCESS SENSING TECHNOLOGIES · TRANSMITTER
Process Sensing Technologies SenzTx
Compact oxygen transmitter offered with zirconia or electrochemical sensing.
Product Overview
Manufacturer: Process Sensing Technologies
Product: SenzTx
Type: Installed measurement transmitter
Range / measured parameters: 0–1,000 ppm and percent O₂ variant ranges
Communication / outputs: 4–20 mA; Modbus options
Engineering and Selection Considerations
Confirm the exact manufacturer’s ordering code, current availability, calibrated measurement range, output protocol, power supply, operating environment and mounting requirements. Technical parameters can change between models, ordered options or product generations.
Official Technical Documentation
Official manufacturer product information →
Independent AirAlign product reference. Source reviewed 13 October 2026. The manufacturer’s technical documents remain authoritative.
PST / Ntron SenzTx — zirconia versus electrochemical O₂ — verified technical selection
Source-specific engineering data for the model and its defined variant. All specifications are maker statements and are not an independent AirAlign test or approval.
| Parameter | Manufacturer specification | Practical limitation |
|---|---|---|
| Sensor technology choices | Zirconia (ZR) or electrochemical (EC) oxygen element | EC is specifically useful for background gases containing hydrocarbons that may be unsuitable for hot zirconia sensing. |
| ZR ranges | 0–1000 ppm, 0–1%, 0–25% or 0–96% O₂ family options | Hardware range/technology is order-code dependent. |
| EC ranges | 0–1000 ppm or 0–25% O₂ per original data sheet | Do not assign the 0–96% ZR range or ZR lifespan to electrochemical version. |
| Accuracy / response | ±2% of reading or 2 ppm O₂ (at calibrated conditions); T90 less than 10 s | Sample flow, pressure and gas composition can change practical response. |
| Supply / communications | 24 V DC, ≤50 mA; 4–20 mA analog plus RS485 Modbus | Use OEM M12 pin assignment and permitted bus loading. |
| Environment by technology | ZR −20…+50 °C; EC 0…+45 °C; 900–1100 mbar absolute | Different technology options have nonidentical ambient conditions. |
| Expected consumable life | 2019 data table: ZR 3–5 years; EC ~1 year, depending on use | Sensor life is not guaranteed indefinitely despite broad family marketing. |
| Body / process connection | IP66 anodized aluminum, 47 mm diameter, 120–140 mm length; KF40 flange, flowthrough or orifice | Specify true sample-line and chamber compatibility. |
System integration and commissioning
The original PST data sheet contrasts non-depleting hot zirconia electrochemistry with consumable EC measurement. It reports ZR longer-term durability while the EC model is more suitable for certain hydrocarbon backgrounds. Do not present any SenzTx as a certified personal oxygen deficiency monitor; it is an OEM industrial process transmitter. At commissioning, verify the 4 mA and 20 mA readings correspond to the chosen ppm or %O₂ scale, identify sensor technology, test Modbus fault paths and document sample pressure/flow control. The older manufacturer revision is clearly labelled; current part-number/certificate and availability need checking for procurement.
Original technical reference: Manufacturer model page, specification or data sheet. Verify the exact purchased suffix and revision, certificates and calibration assumptions.
AirAlign technical editorial enrichment · 25 October 2026. Published specs are not independent laboratory, hazardous-area certificate or active procurement signoff.