HUBA CONTROL · DIFFERENTIAL PRESSURE TRANSMITTERS

Huba Control 450 Differential Pressure Transmitter

Original manufacturer Huba Control Model 450 product photo
Official Huba Control Model 450 example. Connections and specifications vary by ordered variant.

Differential-pressure transmitters for air and non-aggressive gases.

Manufacturer: Huba Control AG.

Ex-proof status: No ATEX/IECEx approval inferred from the ordinary model listing.

Official Source

Huba manufacturer product index / technical documents →

Check measuring range, medium compatibility, output and current approval for the ordered model suffix.

Huba Control Model 450 official current catalog / downloads →

Huba Type 450 — board-mount fast Δp, analogue versus I²C

Huba 450 is a small board-mount ceramic differential-pressure transducer for neutral air and gases. Unlike enclosure-mounted transmitter 699, it is designed for OEM PCB assembly; the manufacturer’s low-pressure variants have substantially different range, burst and interface constraints.

TopicHuba model specificationEngineering and QA consequence
Order rangesRelative/differential family from approximately −1 to +100 mbar (check exact model variant)Do not apply the widest 100 mbar to a low-span 3 mbar unit; specify span and positive/negative port limits.
Dynamics<2 ms sensor response in official model listingVery fast sensor does not make controller flow readings independent of filtering or tubing.
Electrical optionsAmplified temperature-compensated analogue, or digital I²C depending on order; <5 mA supply drawI²C clock ≤100 kHz per Huba guidance; pull-ups ≥1 kΩ and bus ≤200 pF; no standard industrial RS485 or native 4–20mA.
Installation/protectionPCB-mount IP00, polyamide housing, ceramic Al₂O₃ element and silicone diaphragmRequires separately sealed system housing, proper pneumatic tubes and moisture ingress design.
Operating environmentMedium and ambient −20…+85°C; compensated −10…+80°COutside compensation interval accuracy may vary although parts may operate.

Variant-specific specification and approvals

Before installing or certifying the device, verify exact Huba order code, system-pressure limits, process media, signal supply, output scaling, pressure ports, EMC and the final assembly’s applicable certificates. No general-purpose ATEX/IECEx approval is claimed on this page.

OEM source: Huba 450 current technical page.