Posifa PVC6100 MEMS Pirani Vacuum Gauge

Atmosphere-to-medium-vacuum gauge with manufacturer-stated nominal measurement range 10⁻⁴ Torr to atmosphere. Provides 0–10 V analogue signal through FCC-68/RJ45 connector; configurable output curve for some gauge replacement requirements. Industrial applications include freeze drying, mass spectrometry and vacuum furnaces.
Original technical documents
Official manufacturer model / documentation page · PVC6100 manufacturer datasheet (PDF).
Approval/selection: No ATEX/IECEx certification has been established for this model. Manufacturer model codes, connector, fluid compatibility and individual operating parameters must be confirmed from the current datasheet; a model family description is not an order-specific industrial safety approval.
PVC6100 — analog RJ45 gauge, replaceable factory-characterized probe
| Attribute | Original Posifa data | System consequence |
|---|---|---|
| Measurement method | Second-generation MEMS Pirani thermopile sensing element with a low-temperature microheater | Sensitive from 10⁻⁴ Torr to ATM according to published product description. |
| Connection and analogue output | 0–10 V signal through FCC-68/RJ45 connector | Not the I²C bare PCB output of PVC4100; correct wiring and controller voltage scale needed. |
| Field replacement strategy | OEM offers output mapping to emulate other manufacturers’ vacuum gauge curves | Plug-and-play replacement only after connector, mechanical fitting and voltage transfer curve are matched. |
| Contamination resistance | Low heater temperature and sealed microfabricated thermal chamber mitigate filament contamination | Process gas contamination and pressure pulses still need assessment. |
| Use cases | Freeze dryers, mass spectrometers, cryogenic tanks, distillation, vacuum furnaces and load-locks | Validate overpressure, seal, venting and sensor recovery for each process. |
Pressure-range, media and product photo controls
A Pirani gauge’s reading is gas-conductivity dependent. The 10⁻⁴ Torr to ATM envelope is not a universal uncertainty specification. PVC6100 and PVC6800 may share replacement applications but different pressure ranges and composite sensing electronics.
Model source: PVC6100 original Posifa product specifications.