MEMBRAPOR · HYDROGEN SULFIDE (H₂S) SENSORS

Membrapor H2S/CG-10000 Electrochemical Sensor

Illustrative Membrapor compact electrochemical sensor family
Manufacturer-family photograph; individual sensor variants may differ.

Compact electrochemical H₂S sensing family for biogas and high-concentration process monitoring.

Manufacturer: Membrapor

Technical specifications

Target gasH₂S
Range class10,000 ppm
PrincipleElectrochemical
ApplicationBiogas, landfill and process monitoring

Manufacturer documentation

Official product information / technical documents →

Verify exact gas configuration, part number, calibration, response, environmental limits and applicable safety certification before equipment design-in. Product data can change between revisions. AirAlign is independent of the listed manufacturer.

H2S/CG-10000 — high-span electrochemical biogas cell — technical model review

Source-specific manufacturer specifications, test conditions and practical engineering distinctions. The stated model/range is not a blanket safety approval or a guarantee that related part codes perform identically.

ParameterOEM documentationEngineering interpretation
Working principleThree-electrode electrochemical H₂S sensor, compact gas-tight housingRequires an appropriate potentiostat/analog amplifier; no built-in Modbus or ppm screen.
Nominal span0–10,000 ppm H₂SDistinct from common 0–10 or 0–100 ppm personal H₂S alarm cartridges.
Maximum overload20,000 ppm H₂SAn overload survival statement is not a calibrated 20,000-ppm measurement range.
Current sensitivity15 ±5 nA per ppm H₂SHost electronics and calibration determine the usable reading.
Resolution / responseLess than 3 ppm electronics dependent; T90 less than 35 secondsResolution alone makes it unsuitable to infer dependable sub-ppm safety thresholds.
Baseline−60 to +60 ppm equivalent in clean air at 20 °CCheck offset and zero compensation for lower-concentration applications.
Working conditions−40 to +50 °C; 15–90%RH noncondensing; atmospheric pressure ±10%Biogas condensation/pressure and sample conditioning must stay within limits.
Life / storageExpected 2 years in air; 6 months storage in container; 10–33 Ω suggested loadPlan calibration/spares rather than asserting long-term drift-free use.

Installation, electronics and calibration

The model is valuable for discontinuous sampling or biogas analyzer duty because its 10,000-ppm span tolerates concentrations well above normal occupational monitoring ranges. Include sample-line drying/condensate control, a suitable material seal and offset compensation. Membrapor’s own sheet makes performance claims at 20–25 °C, 30–50%RH and 900–1100 mbar and states that the specified resolution depends on the electronics. A bare sensor does not satisfy worker alarm certification merely because the leaflet includes safety/process use cases. Review cross-sensitivity tables before calibrating with blended biogas.

Original manufacturer technical source: Official product specifications and primary data sheet. Exact orderable variant, gas and current certificate must govern procurement or any safety function.

AirAlign manufacturer-specific editorial review · 25 October 2026. Detailed integration information is not an independent calibration or Ex/ATEX/IECEx signoff.