ALPHASENSE / AMETEK · CO₂ SENSOR

Alphasense IRC-A1

A compact Alphasense NDIR carbon-dioxide sensing cell for OEM instruments spanning indoor air quality, safety, combustion and process measurement.

Key Specifications

TECHNOLOGYNDIR · dual-channel pyroelectric detector
MEASURED GASCO₂
CONFIGURABLE RANGES0–5,000 ppm · 0–5% · 0–20% · 0–100% vol.
ACCURACYIAQ ±1% FS · Safety ±1.5% FS · Combustion ±2.5% FS · Process ±4% FS
RESPONSE TIME<40 s at 20 °C
PACKAGEAlphasense A-Series
TEMPERATURE COMPENSATIONIntegrated thermistor output
OPERATING LIFEMTBF >5 years (NDIR cell)

Product Positioning

Alphasense positions IRC-A1 for OEM instruments that require selective CO₂ measurement across ppm and percent-volume ranges. Range-specific performance values should be selected according to the intended IAQ, safety, combustion or process application.

Official Resources

This page uses current Alphasense manufacturer specifications.

View Official Alphasense Product Page

AirAlign is independent of Alphasense and AMETEK. Manufacturer documentation remains authoritative.

Alphasense IRC-A1 — pyroelectric dual-channel NDIR and calibration linearisation — detailed technical comparison and documentation

Manufacturer source data are specific to the selected sensor physics, ordered range and application conditions. Factory product family options do not all apply to a single device.

SpecificationOriginal maker evidenceIntegration/selection qualification
Sensing elementA-Series dual-channel pyroelectric NDIR CO₂ detectorA raw sensor optical/electrical assembly, not a 4–20mA Modbus ready transmitter.
Range optionsIAQ 0–5,000ppm; Safety 0–5%vol; Combustion 0–20%vol; Process 0–100%volThe installed spectral calibration linearisation constants and accuracy depend on ordered CO₂ grade.
Accuracy groupIAQ ±1% full scale, Safety ±1.5% FS, Combustion ±2.5%FS, Process ±4%FS in original OEM tablePercent full scale can be a large absolute error close to zero; do not describe it as percent of reading.
ResponseT90 <40 seconds at 20°C, maker referenceInstalled sample flow/ambient filter changes response.
Onboard compensationThermistor output supports temperature compensation and dual-channel optical referenceA host electronics transfer circuit and calibration remain required.
Original OEM app noteAAN 201-06 publishes separate IRC-A1 IAQ/Safety/Combustion/Industrial linearisation and compensation coefficientsDo not copy coefficients from thermopile IRC-AT or between sensor ranges.
Optical environmental issueOriginal maker notes optional IRC-AF external-light filter when ambient light reaches dust-filter surfaceUncontrolled illumination of a sensor window may affect readings even with onboard compensation.

Design, gas calibration and safety boundaries

For 0–5000ppm IAQ, Alphasense’s original technical material discusses calibration with a 4000ppm reference and normalized linearisation; those steps should be performed by the OEM instrument builder, not assumed pre-solved by the raw A-Series element. Select a certified front end/analogue acquisition that matches the pyroelectric architecture and temperature-compensation coefficients. Manufacturer’s 2022 application note distinguishes IRC-A1 from IRC-AT: copying universal ‘b/c’ coefficients between models can materially bias an apparently precise ppm signal. A part intended for safety instrument development does not itself grant final fixed-detection instrument Ex approval.

Primary manufacturer technical reference: Official exact model and current product documentation. Obtain the current order code, calibration datasheet and any region-specific approval before installation.

AirAlign source-backed product engineering notes · 25 October 2026. Neither an independent calibration certificate nor finished gas detector Ex/SIL approval.